Normalized public record details for FDA recall code Z-0012-2024.

Recall Res Number
Z-0012-2024
Recalling Firm / Manufacturer
FEI Number
3002095335
Recall Status
Open, Classified
Recall Classification
N/A
Date Initiated
2023-07-27
Date Posted
2023-10-03
Date Terminated
N/A
Product Code
DQY
Event Number
92785
Product Quantity
2245 units
Product Description Guider Softip Guide Catheter 6F 90CM MULTI PURPOSE, Catalog number H965100460, cardiac guide catheter

Reason for Recall Stryker Neurovascular has observed that certain lots of Guider Softip XF Guide Catheters appear to have the incorrect tip curve shape. The impacted products were distributed with a tip curve shape which differs from the labeled tip curve shape.

Action Details Stryker issued an URGENT: MEDICAL DEVICE VOLUNTARY Recall notice to it consignees on 07/31/2023. The notice explained the issue, potential risk, and required the following actions be taken: 1. Immediately check your internal inventory for affected devices. 2. Segregate the affected units in a secure location for return to Stryker. 3. Circulate this Recall-Removal notice internally to all interested/affected parties. 4. Maintain awareness of this communication internally until all required actions have been completed within your facility. 5. Inform Stryker if any of the subject devices have been distributed to other organizations. If yes, provide contact details so that Stryker can inform the recipients appropriately.

Distribution Pattern US Nationwide distribution in the state of Indiana.

Code Information / Serial Numbers GTIN 08714729202509, Lot/Batch Numbers: 26059397, 26338716, 25822912, 25822914, 25822915, 25822916, 26059642, 26584704, 26585219, 26745658, 26745659, 26891768, 27177893, 27412261, 27504202, 27568860, 27754498, 27816607, 28100259, 28384661, 28502671, 28737894, 28737895, 26745657, 27177892, 28320873, 28584618, 28795038

Related Establishments / Registrations

Public Record Disclaimer

This page represents an index of public-record FDA device recalls. It does not imply current safety or compliance status of the product. AvaSpecs is a private aggregator and is not affiliated with the FDA or any government agency. This page does not constitute medical, safety, legal, regulatory, or compliance advice. Refer to official government sources for primary safety alerts.