Product Description
Vista Dental Products Sodium Hypochlorite - 6%, 10 - 12cc Pre-filled Syringes, REF 502300
Reason for Recall
The affected lots may be packaged in syringes which include an incorrect bung. The incorrect bung is made of a latex-free isoprene rubber material. After prolonged storage and exposure to contained solutions, the rubber bung may degrade due to chemical interaction with the bleach solution.
Action Details
The firm initiated the recall by letter on 06/27/2019. The letter directed the consignee to quarantine the product and contact the returns department for return instructions.
Distribution Pattern
US, South Africa, Switzerland, Poland, Italy, Singapore, Australia
Code Information / Serial Numbers
Lot/Work Order Numbers: 2018-1348 2018-1585 2018-1586 2018-1852 20182106, UDI (01)10818207020465 20182302, UDI (01)10818207020465 20182458, UDI (01)10818207020465 20190725, UDI (01)10818207020465 20190835, UDI (01)10818207020465
Public Record Disclaimer
This page represents an index of public-record FDA device recalls. It does not imply current safety or compliance status of the product. AvaSpecs is a private aggregator and is not affiliated with the FDA or any government agency. This page does not constitute medical, safety, legal, regulatory, or compliance advice. Refer to official government sources for primary safety alerts.