Product Description
Emission Computed Tomography System. Including Model numbers: 2200967, ASM001080, ASM001083, ASM001092, ASM001099, ASM000085, 10036800109, and Varicam.
Reason for Recall
GE Healthcare received reports of intermittent low ejection fraction (EF) results for MUGA (multi-gated acquisitions) studies on Millennium VG systems. This issue can occasionally lead to distortions in the volume curve in some MUGA studies, and result in calculated EF values lower (never higher) than actual values. A low EF value could lead to a physician decision to stop needed treatment.
Action Details
A letter dated August 5, 2004 was released to customers describing the issue and recommending that users pay special attention to any volume curve which appears to be abnormal or any EF values that appear to be lower than expected, and not to rely soley on this value for diagnosis purposes until they recieve the SW patch. A SW patch will be distributed shortly to all affected systems.
Distribution Pattern
Nationwide throughout the United States and worldwide.
Code Information / Serial Numbers
Each unit has a unique serial code. All serial numbers are being recalled.
Public Record Disclaimer
This page represents an index of public-record FDA device recalls. It does not imply current safety or compliance status of the product. AvaSpecs is a private aggregator and is not affiliated with the FDA or any government agency. This page does not constitute medical, safety, legal, regulatory, or compliance advice. Refer to official government sources for primary safety alerts.