Medical Device Recall: Z-0012-05

FDA Recalls dataset · Firm: General Electric Med Systems LLC · Status: Terminated

Recall Action Class: N/A Date Initiated: 2004-09-03 Recall record compiled from public FDA source records.

Recall Parameters & Description

Normalized public record details for FDA recall code Z-0012-05.

Recall Res Number
Z-0012-05
Recalling Firm / Manufacturer
FEI Number
2126677
Recall Status
Terminated
Recall Classification
N/A
Date Initiated
2004-09-03
Date Posted
2004-10-07
Date Terminated
2006-07-24
Product Code
KPS
Event Number
29988
Product Quantity
853 units
Product Description Emission Computed Tomography System. Including Model numbers: 2200967, ASM001080, ASM001083, ASM001092, ASM001099, ASM000085, 10036800109, and Varicam.

Reason for Recall GE Healthcare received reports of intermittent low ejection fraction (EF) results for MUGA (multi-gated acquisitions) studies on Millennium VG systems. This issue can occasionally lead to distortions in the volume curve in some MUGA studies, and result in calculated EF values lower (never higher) than actual values. A low EF value could lead to a physician decision to stop needed treatment.

Action Details A letter dated August 5, 2004 was released to customers describing the issue and recommending that users pay special attention to any volume curve which appears to be abnormal or any EF values that appear to be lower than expected, and not to rely soley on this value for diagnosis purposes until they recieve the SW patch. A SW patch will be distributed shortly to all affected systems.

Distribution Pattern Nationwide throughout the United States and worldwide.

Code Information / Serial Numbers Each unit has a unique serial code. All serial numbers are being recalled.

Related Establishments / Registrations

Public Record Disclaimer

This page represents an index of public-record FDA device recalls. It does not imply current safety or compliance status of the product. AvaSpecs is a private aggregator and is not affiliated with the FDA or any government agency. This page does not constitute medical, safety, legal, regulatory, or compliance advice. Refer to official government sources for primary safety alerts.