Medical Device Recall: Z-0010-2018

FDA Recalls dataset · Firm: Optovue, Inc. · Status: Terminated

Recall Action Class: N/A Date Initiated: 2017-07-27 Recall record compiled from public FDA source records.

Recall Parameters & Description

Normalized public record details for FDA recall code Z-0010-2018.

Recall Res Number
Z-0010-2018
Recalling Firm / Manufacturer
FEI Number
3005950902
Recall Status
Terminated
Recall Classification
N/A
Date Initiated
2017-07-27
Date Posted
N/A
Date Terminated
2021-04-12
Product Code
HLI
Event Number
77776
Product Quantity
326 devices
Product Description iVue 500 with iScan, D216606. Product non-contact, high resolution tomographic imaging device. It is intended for in vivo imaging, axial cross-sectional and three-dimensional imaging and measurement of anterior and posterior ocular structures, including retina, retinal nerve fiber layer, ganglion cell complex (GCC), optic disc, cornea, and anterior chamber of the eye. The iVue 500 with normative database is a quantitative tool for the comparison of retina, retinal nerve fiber layer, ganglion cell complex, and optic disc measurements to a database of known normal subjects. The iVue 500 with normative database is indicated for use as a device to aid in the diagnosis, documentation, and management of ocular health and diseases in the adult population.

Reason for Recall FDA determined that a 510(k) is needed for the Vault Mapping software (also known as Lens Fitting).

Action Details Optovue sent an Urgent Medical Device Software Field Correction letter dated July 27, 2017, via FedEx mail. Customers were Instructed to discontinue use of the Vault Mapping software, to complete and return the recall response form, and to provide remote access to units to an Optovue representative so that new software can be downloaded remotely. Questions can be addressed by calling Optovue at 1-866-941-9240.

Distribution Pattern US Nationwide Distribution in the states of - AL, AR, AZ, CA, CO, CT, FL, GA, ID, IL, IN, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, VA, WA, WI, & WV.

Code Information / Serial Numbers Cat No. 700-49428-001 (GTIN - N/A), 700-49428-002 (GTIN 00858848006315), 700-49428-003 (GTIN 00858848006018).

Related Establishments / Registrations

Public Record Disclaimer

This page represents an index of public-record FDA device recalls. It does not imply current safety or compliance status of the product. AvaSpecs is a private aggregator and is not affiliated with the FDA or any government agency. This page does not constitute medical, safety, legal, regulatory, or compliance advice. Refer to official government sources for primary safety alerts.