Product Description
Babi.Plus 12.5 cm H2O Pressure Relief Manifold Catalog # 2691 for use with ventilators.
Reason for Recall
Device vented gas below the stated pressure.
Action Details
On June 14 the firm sent communication with the following instructions:
1. Immediately discontinue distribution and quarantine any products with the product codes and lot numbers listed above.
2. Using the provided customer letter and recall acknowledgement form templates, communicate this recall to any of your customers who have received product included within the scope of the recall.
3. Have the customers return any affected product to you, together with a completed Recall Acknowledgement Form, for consolidation and return to Teleflex Medical. In the event that an alternative approach is needed, contact Teleflex Medical Customer Service for more information at 1-866-246-6990.
4. To return recalled products to Teleflex Medical, complete the enclosed Recall Acknowledgement Form and fax it to 1-855-419-8507, Attn: Customer Service or email to recalls@teleflex.com. This will allow us to document the amount of product you have on hand for return. A customer service representative will contact you with a Return Goods Authorization (RGA) Number and will provide instructions for the return of product to Teleflex Medical.
5. Once you have completed returning all of the recalled products from your own inventory, and collecting and consolidating all of the recalled products from your customers, please check the box on the enclosed Recall Acknowledgment Form that indicates that you have completed the recall and fax it to 1-855-419-8507, Attn: Customer Service, or email it to recalls@teleflex.com. This will allow us to document completion of the recall.
6. If you and your customers have no affected stock, please complete the enclosed Recall Acknowledgment Form and fax it to 1-855-419-8507, Attn: Customer Service or email to recalls@teleflex.com. This will allow us to document your receipt of this letter.
Distribution Pattern
US nationwide distribution. AL, CA, CO, CT, FL, GA, HI, IL, IN, KS, KY, LA, MA, MD, ME, MI, MO, MS, NC, NE, NJ, NM, NY, OH, OK, OR, PA, SC, TX, VA, WA and WY
Code Information / Serial Numbers
Catalog # 2691 SKU #: XTF0046 Lot #'s: 180806, 180910, 181029, 181105, 181204, 190225 and 190327
Public Record Disclaimer
This page represents an index of public-record FDA device recalls. It does not imply current safety or compliance status of the product. AvaSpecs is a private aggregator and is not affiliated with the FDA or any government agency. This page does not constitute medical, safety, legal, regulatory, or compliance advice. Refer to official government sources for primary safety alerts.