Product Description
Hyperglycosylated Human Chorionic Gonadotropin (H-hCG) Assay Cartridges (Catalog#62-7039)
Reason for Recall
Data from internal testing indicates that the Nichols Advantage Invasive Trophoblast Antigen Assay Cartridges (ITA)(Lot No 62-500261) and Hyperglycosylated Human Chorionic Gonadotropin Assay Cartridges (H-hCG)(Lot No 62-500262) does not maintain its stability over the 12 month expiration dating of the product, and that stability is assured only through 6 months.
Action Details
Withdrawal letter was sent 08/30/2005 which request customers to stop using the product immediately and discard any remaining materials. A response card was included.
Distribution Pattern
Nationwide and China, Germany, & Australia
Code Information / Serial Numbers
Lot #62-500262
Public Record Disclaimer
This page represents an index of public-record FDA device recalls. It does not imply current safety or compliance status of the product. AvaSpecs is a private aggregator and is not affiliated with the FDA or any government agency. This page does not constitute medical, safety, legal, regulatory, or compliance advice. Refer to official government sources for primary safety alerts.