Medical Device Recall: Z-0006-2026

FDA Recalls dataset · Firm: PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. Veenpluis 4-6 Best Netherlands · Status: Open, Classified

Recall Action Class: N/A Date Initiated: 2025-09-03 Recall record compiled from public FDA source records.

Recall Parameters & Description

Normalized public record details for FDA recall code Z-0006-2026.

Recall Res Number
Z-0006-2026
FEI Number
3002807880
Recall Status
Open, Classified
Recall Classification
N/A
Date Initiated
2025-09-03
Date Posted
2025-10-01
Date Terminated
N/A
Product Code
OWB
Event Number
97502
Product Quantity
501 units (212 US, 289 OUS)
Product Description Allura Xper FD20 Biplane; Model Numbers: (1) 722008, (2) 722013; Associated 510(k)s: K102005, K130638, K130842, K133292, K141979, K162859, K161563;

Reason for Recall The BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the system will be halted, and the system will not start. No user messages are shown before the battery is low on power or depleted.

Action Details On September 3, 2025, URGENT MEDICAL DEVICE CORRECTION letters were sent to customers. Actions to be taken: 1. Circulate this Urgent Medical Device Correction letter to all users of the system so that they are aware of the issue. 2. In case the affected system has been transferred to another organization, please send a copy of this Urgent Medical Device Correction letter to that organization and inform Philips about this transfer through your local Philips representative. 3. Keep this Urgent Medical Device Correction letter with the documentation of the system until Philips corrects your system. Ensure that the letter is in a place likely to be seen/viewed. 4. Complete and return the response form included in this Urgent Medical Device Correction to Philips promptly and no later than 30 days from receipt. 5. If you experience the issue described in this letter, please report it to your local Philips representative. Actions to be taken by recalling firm: 1. Philips will replace the BIOS battery in all affected systems. 2. Philips expects to start the implementation of this action in affected devices by November 2025. Your local Philips representative will contact you to schedule a visit to perform this activity in your system once available. If you need additional information or support concerning these issues, contact your local Philips representative at 1-800-722-9377 (Philips Customer Care Solutions Center).

Distribution Pattern Worldwide distribution - US Nationwide and the countries of Albania, Algeria, Argentina, Australia, Austria, Azerbaijan, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cuba, Curacao, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Finland, France, Georgia, Germany, Greece, Guinea, Honduras, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Korea, South, Kosovo, Latvia, Lebanon, Libya, Lithuania, Macedonia, Malaysia, Malta, Mauritius, Mexico, Mongolia, Morocco, Nepal, Netherlands, New Zealand, Norway, Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Tanzania, Thailand, Trinidad and Tobago, Tunisia, Turkey, T¿rkiye, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Utd.Arab.Emir., Uzbekistan, Venezuela, Vietnam, Yemen.

Code Information / Serial Numbers Model Numbers: (1) 722008, (2) 722013; UDI-DI: (1) N/A, (2) 00884838059061; Serial Numbers: (1) 491, 489, 360, 325, 142, 125, 312, 319, 82, 192, 244, 453, 306, 116, 124, 18270, 524, 106, 224, 463, 197, 264, 100, 294, 246, 357, 212, 287, 154, 459, 71, 112, 472, 483, 465, 419, 38, 454, 2470, 441, 460, 369, 407, 102, 231, 94, 437, 420, 1178, 108, 270, 442, 521, 372, 113, 34, 187, 254, 365, 390, 413, 52, 18, 19, 239, 288, 424, 279, 115, 481, 141, 304, 267, 361, 415, 178, 245, 445, 23, 296, 487, 349, 456, 470, 386, 69, 1236, 37, 59, 455, 217, 446, 272, 160, 406, 350, 167, 247, 416, 435, 48, 447, 179, 410, 196, 458, 525, 480, 476, 370, 327, 169, 121, 259, 266, 128, 186, 290, 385, 253, 450, 359, 391, 283, 323, 205, 425, 346, 388, 376, 150, 394, 123, 339, 513, 337, 93, 268, 200, 87, 110, 46, 542, 530, 334, 122, 468, 250, 482, 63, 305, 213, 546, 273, 433, 475, 398, 31, (2) 505, 313, 217, 492, 34, 410, 153, 360, 290, 74, 83, 241, 404, 126, 244, 106, 383, 346, 461, 307, 1, 442, 435, 293, 253, 297, 46, 289, 638, 342, 69, 267, 402, 214, 86, 603, 149, 424, 140, 254, 255, 266, 475, 519, 518, 303, 194, 115, 150, 129, 496, 599, 441, 124, 151, 179, 550, 377, 575, 631, 562, 178, 330, 309, 601, 327, 308, 264, 78, 268, 141, 175, 357, 85, 40, 216, 637, 422, 174, 109, 72, 218, 228, 511, 389, 152, 479, 561, 262, 626, 61, 164, 497, 170, 379, 84, 125, 564, 573, 426, 556, 42, 100, 495, 500, 96, 515, 105, 428, 427, 531, 375, 316, 278, 347, 462, 249, 205, 246, 43, 639, 304, 634, 147, 15, 567, 168, 607, 283, 52, 171, 66, 36, 585, 271, 117, 443, 107, 343, 223, 385, 165, 453, 370, 22, 367, 108, 352, 291, 71, 259, 455, 468, 368, 239, 315, 325, 532, 598, 440, 172, 193, 301, 415, 540, 62, 299, 184, 185, 187, 188, 437, 189, 183, 186, 190, 413, 191, 560, 490, 300, 295, 81, 448, 444, 400, 632, 110, 111, 624, 65, 633, 630, 361, 549, 417, 449, 506, 177, 629, 48, 476, 527, 276, 247, 156, 499, 50, 167, 132, 302, 618, 608, 265, 70, 457, 547, 312, 321, 366, 390, 544, 236, 157, 243, 144, 525, 238, 570, 161, 349, 359, 454, 613, 623, 222, 335, 445, 538, 416, 628, 619, 47, 206, 620, 384, 526, 332, 358, 509, 470, 378, 296, 227, 31, 58, 589, 616, 447, 530, 232, 324, 311, 176, 181, 554, 134, 163, 104, 403, 478, 459, 17, 2266, 528, 318, 29, 28, 237, 559, 372, 256, 374, 487, 405, 380, 434, 130, 473, 551, 557, 294, 412, 128, 388, 207, 258, 429, 95, 411, 353, 542, 533, 622, 182, 574, 139, 137, 250, 563, 502, 450, 436, 406, 524, 354, 510, 395, 458, 261, 503, 348, 595, 586, 535, 569, 425, 334, 344, 355, 465, 345, 501, 522, 64, 396, 408, 208, 555, 394, 57, 493, 201;

Related Establishments / Registrations

Public Record Disclaimer

This page represents an index of public-record FDA device recalls. It does not imply current safety or compliance status of the product. AvaSpecs is a private aggregator and is not affiliated with the FDA or any government agency. This page does not constitute medical, safety, legal, regulatory, or compliance advice. Refer to official government sources for primary safety alerts.