Product Description
Z-800 Infusion System, Model Numbers Z-800, Z-800F, Z-800W, Z-800WF;
Software Version: Z-800 6.1.01 and 6-1.07z; Z-800F 4.1.02 and 4.1.08z; Z-800W 3.1.32 and 3.1.64z, Z-800WF 3.1.32 and 3.1.64z
Reason for Recall
There is a defect in the air-in-line software algorithm.
Action Details
An "URGENT: MEDICAL DEVICE CORRECTION" notification letter dated 9/13/24 was sent to customers.
Mitigation
Zyno Medical LLC is mitigating this software issue by conducting a voluntary medical device correction to remove the faulty software and correctly install software that is free of this defect.
Recommended Actions
1. Actions to Address the Issue During Use
Note that only the devices listed in this letter are affected. No other products are involved in this action.
Discontinue use of the affected device and proceed to the instructions for return of the device in Recommended Action below.
2. Actions to Correct the Issue
Complete the attached Zyno Medical Z-800, Z-800F, Z-800W, and Z-800WF Infusion Pump Air-in-Line Software Issue Response Verification Form. This will include checking your inventory for the affected products, verifying serial numbers, confirming product location, reporting any transferred product, and signing and dating the completed form.
A representative from Zyno Medical s partner company, Intuvie LLC, will be in contact with you to coordinate instructions on exchanging devices to remediate this action.
Pass this notice to the appropriate personnel who need to be informed if the identified devices have been transferred to another location.
If you have any questions or would like assistance, please contact your local Business Development Manager.
Contact Information: Customerservice@intuvie.com
***Update 1/3/2025***Additional consignees were sent URGENT: CLASS 1 RECALL letters due to additional affected units identified. Recommended Actions
1. Actions to Address the Issue During Use
Note that only the devices listed in this letter are affected. No other products are involved in this action.
Discontinue use of the affected device and proceed to the instructions for return of the device in Recommended Action below.
2. Actions to Correct the Issue
Complete the attached Zyno Medical Z-800, Z-800F, Z-800W, and Z-800WF I
Distribution Pattern
US Nationwide.
Code Information / Serial Numbers
UDI-DI:
Z-800 - 00814371020006
Z-800W - 00814371020020
Z-800F - 00814371020013
Z-800WF - 00814371020037
Lot Numbers:
190722360
20110921-SH
20111005-SH
20111102-SH
20111212-SH
20120105-SH
20120222-SH
20120501-SH
20120612-SH
20120719-SH
20120809-SH
20190809-SH
20120910-SH-1
20121213-sh
20130123-SH
20130227-SH
20120417-SH
20130511-SH
20130517-SH
20130609-SH
20130701-SH
20130812-SH
20131024-SH
Public Record Disclaimer
This page represents an index of public-record FDA device recalls. It does not imply current safety or compliance status of the product. AvaSpecs is a private aggregator and is not affiliated with the FDA or any government agency. This page does not constitute medical, safety, legal, regulatory, or compliance advice. Refer to official government sources for primary safety alerts.