Medical Device Recall: Z-0005-2019

FDA Recalls dataset · Firm: Varian Medical Systems · Status: Terminated

Recall Action Class: N/A Date Initiated: 2018-08-20 Recall record compiled from public FDA source records.

Recall Parameters & Description

Normalized public record details for FDA recall code Z-0005-2019.

Recall Res Number
Z-0005-2019
Recalling Firm / Manufacturer
FEI Number
3006583922
Recall Status
Terminated
Recall Classification
N/A
Date Initiated
2018-08-20
Date Posted
N/A
Date Terminated
2020-08-28
Product Code
IYE
Event Number
81010
Product Quantity
4
Product Description VitalBeam Radiotherapy Delivery System Version 2.5 Product Usage: The VitalBeam delivery systems are intended to provide precision radiotherapy and stereotactic radiosurgery for lesions, tumors, and conditions anywhere in the body where radiation treatment is indicated.

Reason for Recall Reports have been received of an anomaly that can result in a treatment without intended gating (respiratory tracking/monitoring). This issue occurs when a patient planned with gating is treated on more than one system.

Action Details On 08/20/2018 the firm sent an URGENT MEDICAL DEVICE CORRECTION FIELD SAFETY NOTICE to customer via courier service (US and Canada via FedEx on 8/23/2018, and Rest of World via third party courier on 8/24/2018) informing them the firm had received reports of an anomaly that can result in a treatment without intended gating, and that this issue occurs when a patient planned with gating is treated on more than one system. Customers were advised, for treatment plans with intended gating, the firm recommends users monitor and pay attention to the system status message. When a patient with a gated treatment plan is moved to a second system, users must verify that intending gating was not overridden. If gating is overridden, users should reselect it from the Tools menu. The customer were informed that they will be contacted by a Customer Service Representative to schedule upgrade to the correction when available and if they have any questions regarding selecting gating from the Tools menu to contact customer support. Help Desk Contact Information: Phone: +1.888.827.4265 Email: support @varian.com In addition, the URGENT MEDICAL DEVICE CORRECTION FIELD SAFETY NOTICE, informs the customers to: 1) Please retain a copy of this document along with your most current product labeling. 2) Advise the appropriate personnel working in your radiotherapy department of the content of this letter. 3) For future reference, this document will be posted to the Varian Medical Systems customer support website: http://www.MyVarian.com. and In order to satisfy regulatory requirements, we request that you complete the attached Recall Return Response form provided. Kindly return the completed form to returnresponse@varian.com.

Distribution Pattern Worldwide Distribution - US nationwide in the states of AL, AR, AZ, CA, CO, FL, Guan, IL IN, KS, KY, LA, MA, MD, MI, MO, NH, NJ, NY, OH, OK, PA, Puerto Rico, TN, TX, VA, WI and the countries of : Algeria, Australia, Azerbaijan, Belgium, Brazil, Bulgaria, Canada, China, Croatia, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Georgia, Germany, India, Ireland, Israel, Italy, Japan, Kazakhstan, Lebanon, Morocco, Nepal, The Netherlands, New Zealand, Norway, Panama, Poland, Reunion Russia, Saudi Arabia, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey and United Kingdom

Code Information / Serial Numbers VitalBeam Version 2.5MR1 Systems (software 02.05.13.05/02.05.13.07): H192482 VITAL BEAM - 321274492 H192521 VITAL BEAM H192743 VITAL BEAM H192816 VITAL BEAM  Varian Product Code family H19, which includes TrueBeam, TrueBeam STx, VitalBeam and EDGE.

Related Establishments / Registrations

Public Record Disclaimer

This page represents an index of public-record FDA device recalls. It does not imply current safety or compliance status of the product. AvaSpecs is a private aggregator and is not affiliated with the FDA or any government agency. This page does not constitute medical, safety, legal, regulatory, or compliance advice. Refer to official government sources for primary safety alerts.