Medical Device Recall: Z-0004-2022

FDA Recalls dataset · Firm: Abbott Molecular, Inc. · Status: Terminated

Recall Action Class: N/A Date Initiated: 2021-09-02 Recall record compiled from public FDA source records.

Recall Parameters & Description

Normalized public record details for FDA recall code Z-0004-2022.

Recall Res Number
Z-0004-2022
Recalling Firm / Manufacturer
FEI Number
3005248192
Recall Status
Terminated
Recall Classification
N/A
Date Initiated
2021-09-02
Date Posted
2021-10-14
Date Terminated
2024-07-23
Product Code
QJR
Event Number
88626
Product Quantity
60563 units
Product Description Alinity m SARS-CoV-2 Amp Kit (Emergency Use Authorization, US Assay) and Alinity m SARS-CoV-2 Application Specification File. in vitro diagnostic

Reason for Recall There is a potential for false positive results.

Action Details The firm issued an Urgent Field Safety Notice/Field Correction Recall 09/02/2021 to all impacted customers. Per the customer letter, all positives are to be considered presumptive until the Alinity m SARS-CoV-2 Application Specification file version 5 or the updated Alinity m Resp-4-Plex Application Specification file is implemented at the customer site. On 03/30/2022, Abbott Molecular issued a follow-up Urgent Field Safety Notice/Field Correction Recall letter to all customers who ordered Alinity m SARS-CoV-2 AMP Kit Emergency Use Authorization (EUA), List Number 09N78-095 and/or Alinity m Resp-4-Plex AMP Kit EUA, List Number 09N79-096 used with the Alinity m System. Also to customers who ordered Alinity m SARS-CoV-2 AMP Kit (CE), List Number 09N78-090 or 09N78-091 and/or Alinity m Resp-4-Plex AMP Kit (CE), List Number 09N79-090 used with the Alinity m System. The Customer Letter explains Abbott has updated the existing Application Specification files for Alinity m SARS-CoV-2 and Resp-4-Plex to further reduce the potential for carryover. FDA has not yet authorized the Alinity m SARS-CoV-2 Application Specification file version 6 update or the Alinity m Resp-4-Plex Application Specification file version 5 update. Implementation of the modification for Alinity m SARS-CoV-2 and Resp-4-Plex will occur when the update to the Application Specification file has been authorized. A Molecular Diagnostics Abbott Representative will be contacting you once the updated Application Specification files are available. If you have any questions, please contact your local Molecular Diagnostics Abbott representative or call Abbott Technical Support at 1-800-553-7042 Option 2.

Distribution Pattern Worldwide Distribution. US Nationwide, AUSTRALIA, AUSTRIA, BELGIUM, BRAZIL, CANADA, COLOMBIA, CZECH REPUBLIC, ESTONIA, FINLAND, FRANCE, GERMANY, HONG KONG, IRELAND, ISRAEL, ITALY, JAPAN, MALAYSIA, MONTENEGRO, NETHERLANDS, NEW ZEALAND, NORWAY, POLAND, PORTUGAL, ROMANIA, SAUDI ARABIA, SINGAPORE, SLOVENIA, SOUTH AFRICA, SOUTH KOREA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, THAILAND, UNITED KINGDOM, and VIETNAM.

Code Information / Serial Numbers List Number: 09N78-095, UDI 00884999049222; Lot/Serial, Expiration: 511810, 09/18/2021; 512026, 09/21/2021; 512900, 10/14/2021; 513146, 10/14/2021; 513583, 11/18/2021; 513959, 11/23/2021; 514176, 11/23/2021; 514216, 12/01/2021; 514819, 12/17/2021; 515052, 01/05/2022; 515653, 01/28/2022; 515949, 02/05/2022; 516085, 02/11/2022; 516611, 02/22/2022; 517139, 03/01/2022; 517355, 03/08/2022; 517999, 03/19/2022; 518266, 03/22/2022; 518435, 03/22/2022; 522689, 09/14/2022; 518670, 04/01/2022; 522691, 09/14/2022; 519124, 05/03/2022; 523309, 09/13/2022; 519583, 05/11/2022; 523468, 09/23/2022; 519785, 05/18/2022; 523621, 10/01/2022; 519968, 05/25/2022; 523946, 10/08/2022; 520216, 06/08/2022; 524203, 10/21/2022; 520699, 06/30/2022; 524438, 10/25/2022; 521106, 07/13/2022; 525862, 12/16/2022; 521654, 07/23/2022; 526394, 12/30/2022; 521842, 07/27/2022; 526484, 01/14/2023; 521857, 07/27/2022; 527399, 01/28/2023; 527400, 01/28/2023; 527540, 02/04/2023; 528026, 02/18/2023; List File 09N78-03E, UDI (01)00884999050211(240)09N78-03E(8012)5.oo

Related Establishments / Registrations

Public Record Disclaimer

This page represents an index of public-record FDA device recalls. It does not imply current safety or compliance status of the product. AvaSpecs is a private aggregator and is not affiliated with the FDA or any government agency. This page does not constitute medical, safety, legal, regulatory, or compliance advice. Refer to official government sources for primary safety alerts.

Account Required Full available specs will require a free AvaSpecs account.