Medical Device Recall: Z-0004-2017

FDA Recalls dataset · Firm: Leonhard Lang Medizintechnik GmbH Archenweg 56 Innsbruck Austria · Status: Terminated

Recall Action Class: N/A Date Initiated: 2016-09-01 Recall record compiled from public FDA source records.

Recall Parameters & Description

Normalized public record details for FDA recall code Z-0004-2017.

Recall Res Number
Z-0004-2017
FEI Number
6736
Recall Status
Terminated
Recall Classification
N/A
Date Initiated
2016-09-01
Date Posted
2016-10-14
Date Terminated
2019-03-26
Product Code
MKJ
Event Number
75181
Product Quantity
11,110 ( US 8,040; OUS 3,070)
Product Description Skintact Electrodes for Defibrillation, DF29N. Product Usage: Multifunction electrode for external defibrillation, pacing, cardioversion, and monitoring. The device is non-sterile and for single use only.

Reason for Recall There is a risk that defibrillation electrodes model DF29N will be connected with the defibrillator Welch Allyn AED 10 only with delay or not at all. This may cause a situation in which a patient, who is in a life threatening condition and requires a defibrillation shock, cannot be treated in good time.

Action Details Consignees was sent via e-mail a Leonhard Lang "Important Safety Notice" dated September 1, 2016. The letter described the product being recalled, Description of the defect, Actions and time frame of the recall, & Compensation for the recalled electrodes. Advised consignees to inform all users within their organizations of the recall; and, secure and destroy the unused electrodes in their inventory. Confirmation of destruction can be done by completing and returning the "Confirmation of Destruction / Consumption" form by October 14, 2016. If the product was further distributed, they were to forward a copy of the Safety Notice. For questions contact sales staff.

Distribution Pattern Worldwide Distribution - US, to the state of Florida; and, the countries of South Africa, Germany, Slovenia, France, United Arab Emirates, Great Britain, Italy, Israel, Serbia, Thailand, Thailand, South Africa, Lebanon, and Poland.

Code Information / Serial Numbers US: 60602-0774; 60502-0779; 60308-0771; 60114-0773; 51023-0775; 50904-0777; 50403-0778; 50130-0777; 41023-0771; 41008-0778; 40730-0778; 40618-0778; and, 40130-0776. Outside US: 60725-0774; 60620-0776; 60602-0774; 60502-0779; 60308-0771; 60114-0773; 51023-0775; 50904-0777; 50403-0778; 40827-0777; 40730-0778; 40618-0778; and, 40130-0776.

Related Establishments / Registrations

Public Record Disclaimer

This page represents an index of public-record FDA device recalls. It does not imply current safety or compliance status of the product. AvaSpecs is a private aggregator and is not affiliated with the FDA or any government agency. This page does not constitute medical, safety, legal, regulatory, or compliance advice. Refer to official government sources for primary safety alerts.

Account Required Full available specs will require a free AvaSpecs account.