Medical Device Recall: Z-0003-2015

FDA Recalls dataset · Firm: Siemens Medical Solutions USA, Inc · Status: Terminated

Recall Action Class: N/A Date Initiated: 2014-09-02 Recall record compiled from public FDA source records.

Recall Parameters & Description

Normalized public record details for FDA recall code Z-0003-2015.

Recall Res Number
Z-0003-2015
Recalling Firm / Manufacturer
FEI Number
2240869
Recall Status
Terminated
Recall Classification
N/A
Date Initiated
2014-09-02
Date Posted
2014-10-02
Date Terminated
2015-02-24
Product Code
KPR
Event Number
69258
Product Quantity
1
Product Description Siemens AXIOM Luminos dRF Max systems The Axiom Luminos dRF is intended to be used as a universal diagnostic imaging system for radiographic and fluoroscopic studies. Using a digital flat detector, it can perform a range of applications including general R/F, angiography and pediatric examinations. The Axiom Luminos dRF is a device intended to visualize anatomical structures by converting a pattern of X-ray into a visible image.

Reason for Recall Two malfunctions posing potential risk to patients were identified with Siemens Ysio Max, AXIOM Luminos dRF Max, Uroskop Omnia Max, and Luminos Agile Max systems with a specific set of serial numbers that will cause lost images due to an automatic Radiology Information System (RIS) worklist update, thus causing examinations to be repeated. Image label may also be displayed incorrectly after images

Action Details Siemens sent a Safety Advisory Notice, dated September 2, 2014, to all affected customers. The Notice identified the product, the problem, and the action to be taken by the customer. Customers were instructed to disable the automatic RIS worklist update and perform this procedure manually. Customers should disable the Measurement Field Display and Collimator Display by setting the display time for both setting to 0 sec (if editable). Customers should forward the safety notice to the new owner and inform Siemens of the new owner where possible. Customers with questions should call 610-219-2119 or 610-219-4834. For questions regarding this recall call 610-219-6300.

Distribution Pattern Nationwide Distribution - US including UT, MO, CO, MA, MN, MS, OH, SC, MI, and NY.

Code Information / Serial Numbers Siemens AXIOM Luminos dRF Max system material #s: 10762471 with serial number: 5037.

Related Establishments / Registrations

Public Record Disclaimer

This page represents an index of public-record FDA device recalls. It does not imply current safety or compliance status of the product. AvaSpecs is a private aggregator and is not affiliated with the FDA or any government agency. This page does not constitute medical, safety, legal, regulatory, or compliance advice. Refer to official government sources for primary safety alerts.