Medical Device Recall: Z-0002-2023

FDA Recalls dataset · Firm: Fresenius Kabi USA, LLC · Status: Terminated

Recall Action Class: N/A Date Initiated: 2022-09-08 Recall record compiled from public FDA source records.

Recall Parameters & Description

Normalized public record details for FDA recall code Z-0002-2023.

Recall Res Number
Z-0002-2023
Recalling Firm / Manufacturer
FEI Number
3014732157
Recall Status
Terminated
Recall Classification
N/A
Date Initiated
2022-09-08
Date Posted
2022-10-06
Date Terminated
2024-09-10
Product Code
PHC
Event Number
90845
Product Quantity
4 instances
Product Description Ivenix Infusion System (IIS), LVP Software; LVP-SW-0004 version 5.2.0

Reason for Recall The display screen may become frozen and unresponsive to user input, triggering an alarm that is evidenced by a buzzer sound and flashing red LEDs as a secondary notification. However, as the screen goes blank it does not indicate an alarm condition or respond. This issue leads to temporary cessation of flow of a medication or fluid, requiring immediate intervention by the clinician.

Action Details On September 8, 2022, the company notified affected customers via phone call and emailed letter. Customers were instructed to do the following: 1. If a high priority Pump Problem alarm occurs, the LVP should be powered down by holding the power button for 10 seconds then releasing the power button so that the screen goes dark, then pressing the power button again to re-start the pump. Alternatively, if you have another LVP available, you may use it to resume infusion. 2. Inform potential users of the product within your organization of this notification. If your facility further distributes or transfers products amongst satellite sites or other locations, please disseminate this information accordingly. Fresenius Kabi has updated the software to rectify the condition. The updated software is currently available. Please schedule a service call at your earliest convenience to have it installed. Customer support: Ivenix_support@fresenius-kabi.com; 1-978-775-8100

Distribution Pattern Domestic distribution to NJ and WI.

Code Information / Serial Numbers UDI-DI: 00811505030122 version 5.2.0

Related Establishments / Registrations

Public Record Disclaimer

This page represents an index of public-record FDA device recalls. It does not imply current safety or compliance status of the product. AvaSpecs is a private aggregator and is not affiliated with the FDA or any government agency. This page does not constitute medical, safety, legal, regulatory, or compliance advice. Refer to official government sources for primary safety alerts.