Medical Device Recall: Z-0001-2025

FDA Recalls dataset · Firm: Medtronic MiniMed, Inc. · Status: Open, Classified

Recall Action Class: N/A Date Initiated: 2024-07-31 Recall record compiled from public FDA source records.

Recall Parameters & Description

Normalized public record details for FDA recall code Z-0001-2025.

Recall Res Number
Z-0001-2025
Recalling Firm / Manufacturer
FEI Number
3003166194
Recall Status
Open, Classified
Recall Classification
N/A
Date Initiated
2024-07-31
Date Posted
2024-10-03
Date Terminated
N/A
Product Code
OZO
Event Number
95121
Product Quantity
24,595
Product Description MiniMed 630G Insulin Pump, REF: MMT-1515, MMT-1714, MMT-1715, MMT-1754, MMT-1755; MiniMed 700G Insulin Pump, REF: MMT-1800, MMT-1801, MMT-1805, MMT-1850, MMT-1851

Reason for Recall Insulin pumps that have been dropped, bumped, or experienced physical impact may have damage to internal electrical components, which may cause reduced pump battery life. Since the defect reduces the battery life overall, it shortens the time between low battery alarms and increases the frequency with which the user must insert a new battery. Further, the defect may also deliver the low battery alarm when there is significantly less time before the battery completely runs out, from the expected up to 10-hour buffer the low alert is supposed to signal, leading to device powering down and stopping insulin delivery unexpectedly. Per the firm, a subset of pumps with the defect showed the low alert alarm was delivered ~2.5 hours before the device shut off, however, patient reports indicate the time could be even shorter. Replacing the battery will not resolve this issue and the short battery life and the truncated low battery alert lead time will continue to occur. Early battery depletion could result in power loss and insulin delivery could be interrupted, resulting in under delivery of insulin potentially leading to hyperglycemia and/or DKA.

Action Details On 7/31/24, Safety Alert notices were sent to customers, healthcare professionals, distributors who were told the following: 1) Always pay attention to alerts and alarms displayed on your pump. - Low Battery Pump alert will display when your pump has up to 10 hours of battery life left. - Replace Battery alert will display when the pump has less than 30 minutes of battery life left. If the battery is not replaced within 10 minutes, a siren will sound and repeat once every minute until the Replace Battery Now alarm is displayed. At that time, the pump will stop insulin delivery. If the battery is not replaced within 10 minutes of this alarm, a siren will sound, and the pump will shut down. 2) If your pump has been dropped, bumped, or has experienced physical impact, pay special attention to any pump alerts and alarms, including low battery alerts, as they may occur earlier than expected. 3) Pump's battery level is on the status bar located on the pump home screen. 4) Refer to the pump user guide for instructions on how to replace the battery. 5) If you notice any significant changes in battery life or need additional troubleshooting assistance, the firm's Technical Support line for further help. 6) Always have extra new AA lithium or alkaline batteries or fully charged NiMH batteries available, along with your other emergency kit supplies. 7) Complete and return the confirmation form by visiting https://info.medtronicdiabetes.com/battery-statusacknowledgement In addition, distribution partners and service providers were provided with a patient letter they could use to inform customers/pump users. If you have further questions or need assistance, please call the firm's 24-Hour Technical Support line at 800-646-4633, option 1.

Distribution Pattern US: CT, MI, PA, WA, IA, NY, ND, AZ, TX, OH, NC, AL, MN, IN, NJ, KY, UT, CA, FL, VA, MS, NM, NV, TN, GA, MA, NH, OK, VT, IL, ME, SC, LA, WY, RI, SD, KS, WI, MD, CO, DE, AR, AK, ID, MO, NE, WV, MT, OR, DC, HI, VI, PR. OUS: Worldwide

Code Information / Serial Numbers REF/UDI-DI(GTIN): MMT-1714/00643169782396, 00763000253059; MMT-1715/00643169873841, 00763000190446, 00643169752726, 643169873834, 00763000317591, 00763000316655, 763000367053, 763000316655, 76300031665503, 763000190439, 763000316631, 763000166519, 763000072520, 76300031665501, 763000256807, 763000090197, 643169656840, 763000316686, 00643169656840, 00763000090197, 00763000190439, 00763000316631; MMT-1755/643169752726, 643169873841, 763000190446, 763000101527, 763000317591, 00763000317591, 763000367091, 763000179632, 763000317621, 643169878648, 763000157043, 76300031759103, 00643169752726, 76300031759101, 76300031762103, 76300031762101, 00763000101527, 00763000190446; MMT-1801/00763000531157; MMT-1805/00763000561680; MMT-1850/00763000552541; MMT-1851/00763000854737, 00763000541323. All serial numbers.

Related Establishments / Registrations

Public Record Disclaimer

This page represents an index of public-record FDA device recalls. It does not imply current safety or compliance status of the product. AvaSpecs is a private aggregator and is not affiliated with the FDA or any government agency. This page does not constitute medical, safety, legal, regulatory, or compliance advice. Refer to official government sources for primary safety alerts.