Product Description
Endo GIA Auto Suture Gray Universal Articulating Loading Unit 30 mm - 2.0 mm, Item Code 030450 - Product Usage: The Endo GIA universal staplers have applications in abdominal, gynecologic, pediatric and thoracic surgery for resection, transection and creation of anastomosis.
Reason for Recall
The device staples may not properly form upon application preventing adequate hemostasis. Use of a product with this assembly error may result in incomplete staple formation.
Action Details
Urgent Medical Device Recall notification letters dated 8/12/20 were sent to customers.
Required Actions:
1.Please immediately quarantine and discontinue use of affected item codes with associated lot numbers listedon Attachment A.
2.Please return affected product as indicated below. All unused products from the affected item codes and associated lot numbers must be returned.
3.If you have distributed the Covidien Endo GIA" Auto Suture" universal articulating loading units listed on Attachment A, please promptly forward the information from this letter to those recipients.
4.Complete the Recalled Product Return Form even if you do not have inventory.
We regret any inconvenience this may cause. We are committed to patient safety and appreciate your prompt attention to this matter. If you have any questions regarding this communication, please contact your Medtronic representative or Customer Service at 800-962-9888, option 2.
Distribution Pattern
Worldwide distribution - US Nationwide and Global Distribution.
Code Information / Serial Numbers
Lot Numbers: T0A108X T7H003X T8D008X T8K108X T9G060X T0E051X T7L023X T8H047X T8M079X T7G003X T8D007X T8K047X T9F132X
Public Record Disclaimer
This page represents an index of public-record FDA device recalls. It does not imply current safety or compliance status of the product. AvaSpecs is a private aggregator and is not affiliated with the FDA or any government agency. This page does not constitute medical, safety, legal, regulatory, or compliance advice. Refer to official government sources for primary safety alerts.