Medical Device Recall: Z-0001-2009

FDA Recalls dataset · Firm: Philips Medical Systems North America Co. Phillips · Status: Terminated

Recall Action Class: N/A Date Initiated: 2008-04-10 Recall record compiled from public FDA source records.

Recall Parameters & Description

Normalized public record details for FDA recall code Z-0001-2009.

Recall Res Number
Z-0001-2009
FEI Number
1000524572
Recall Status
Terminated
Recall Classification
N/A
Date Initiated
2008-04-10
Date Posted
2009-01-26
Date Terminated
2010-10-29
Product Code
JAA
Event Number
50138
Product Quantity
19 units in the US.
Product Description NeuViz Dual Multi-Slice Computed Tomography (CT) Scanner System (software). Whole body x-ray computed tomography scanner which features a continuously rotating tube-detector system and functions according to the fan beam principle. The system software is an interactive program used for x-ray scan control, image reconstruction, and image archive/evaluation. It is intended to produce cross-section images of head and body by computer reconstruction of x-ray transmission data taken at different angles. The function of Time Scan is intended to use in contrast scan include single phase and multi-phase scan, that is according to different density of contrast in vessels or organs between different phase to get necessary information to support doctor's diagnosis.

Reason for Recall An issue associated with the "Timed Scan" application was found. The system does not wait for the programmed time interval and starts the x-ray after selection of the scan button when certain criteria is met.

Action Details A letter was issued to customers prior to the release of the field correction during June 2008. The system software will be updated to correct the non-compliances. For information or support concerning this issue, please contact the Neusoft Medical Systems Service Support Department (helpdesk@pnms.neusoft.com) or contact Philips Medical Systems North America Co. Phillips at 425-487-7602.

Distribution Pattern Nationwide Distribution --- USA including states of CO, FL, GA, IL, KY, NC, TN and TX.

Code Information / Serial Numbers Serial numbers: 400562, 400563, 400626, NDH009EI, NDH011EI, NDH012EI, NDH013EI, NDH016EI, NDH017EI, NDH025EI, NDH026EI, NDH027EI, NDH028EI, NDH030EI, NDH031EI, NDH034EI, NDHR080001, NDHR080002 and NDHR080003.

Related Establishments / Registrations

Public Record Disclaimer

This page represents an index of public-record FDA device recalls. It does not imply current safety or compliance status of the product. AvaSpecs is a private aggregator and is not affiliated with the FDA or any government agency. This page does not constitute medical, safety, legal, regulatory, or compliance advice. Refer to official government sources for primary safety alerts.