Product Description
GE 4 inch Anterior Cervical Post, GE Part Number 1006385 or 1006385-NAV, GE Healthcare OEC Medical Systems (an accessory used with the InstaTrak 3500 plus and 9800 C-Arm Navigation systems)
Reason for Recall
Weld defect- the weld between the post body and the screw was not properly formed and can fail, even when minimal force is used.
Action Details
Consignees were notified by letter on 05/31/2006
Distribution Pattern
Nationwide, including one VA facility in PA. No military distribution. Foreign distribution to Canada, China, Italy, Japan, Spain.
Code Information / Serial Numbers
ALL LOT NUMBERS -GE Part number 1006385 or 1006385-NAV
Public Record Disclaimer
This page represents an index of public-record FDA device recalls. It does not imply current safety or compliance status of the product. AvaSpecs is a private aggregator and is not affiliated with the FDA or any government agency. This page does not constitute medical, safety, legal, regulatory, or compliance advice. Refer to official government sources for primary safety alerts.