| Z-0039-2010 |
Infant Manual Pulmonary Resuscitator with reservoir bag, face mask & Gas sampling elbow. Product ID # 5801, 12-8713, 732-E, and 2206.
Packaged under the name of Dynarex, Evergreen, Mada and Hospitak.
Recalled codes may be identified by examining the patient valve housing immediately below the right-angle exhalation port, where the retention ring should be visible. MPR units with a clear or transparent ring, as well as those where no ring can be seen, should be returned. MPR units with a blue retention ring are not affected and do not need to be returned.
Reason: In some Manual Pulmonary Resuscitator units the duckbill valve and retention ring inside the patient valve housing may become dislodged and fall into the MPR bag/body. If this occurs, the MPR may lose its ability to generate the positive pressure necessary necessary to ventilate the patient.
|
2009-10-06 |
Terminated
|
View Details
|
| Z-0040-2010 |
Pediatric Manual Pulmonary Resuscitator with reservoir bag, face mask & Gas sampling elbow. Product ID #s: 5804; 12-8709, 727-E; 727MM, 2203.
Packaged under the name of Dynarex, Evergreen, Mada and Hospitak.
Reason: In some Manual Pulmonary Resuscitator units the duckbill valve and retention ring inside the patient valve housing may become dislodged and fall into the MPR bag/body. If this occurs, the MPR may lose its ability to generate the positive pressure necessary necessary to ventilate the patient.
|
2009-10-06 |
Terminated
|
View Details
|
| Z-0041-2010 |
Adult Manual Pulmonary Resuscitator with reservoir tube, face mask & Gas Sampling elbow. Product ID # 5807; 12-8703, 723-E (Foreign Distribution only), 723MM.
Packaged under the name of Dynarex and Evergreen.
Reason: In some Manual Pulmonary Resuscitator units the duckbill valve and retention ring inside the patient valve housing may become dislodged and fall into the MPR bag/body. If this occurs, the MPR may lose its ability to generate the positive pressure necessary necessary to ventilate the patient.
|
2009-10-06 |
Terminated
|
View Details
|
| Z-0042-2010 |
Adult Manual Pulmonary Resuscitator with reservoir bag, face mask & Gas sampling elbow. Product ID #s 5808, 12-8704; 724-E; 724MM, 32612.
Packaged under the name of Dynarex, Evergreen, Hospitak and Select Medical Products.
Reason: In some Manual Pulmonary Resuscitator units the duckbill valve and retention ring inside the patient valve housing may become dislodged and fall into the MPR bag/body. If this occurs, the MPR may lose its ability to generate the positive pressure necessary necessary to ventilate the patient.
|
2009-10-06 |
Terminated
|
View Details
|
| Z-0043-2010 |
Adult Manual Pulmonary Resuscitator with reservoir tube, face mask & standard elbow. Product ID #: 12-8702; 778500; 721-E, 778500; 2200; 2201, 86-721E, 40-116722 (foreign distributiuon only), 721MM.
Packaged under the name of Evergreen, Hospitak, Kentron, McKesson and Mada.
Reason: In some Manual Pulmonary Resuscitator units the duckbill valve and retention ring inside the patient valve housing may become dislodged and fall into the MPR bag/body. If this occurs, the MPR may lose its ability to generate the positive pressure necessary necessary to ventilate the patient.
|
2009-10-06 |
Terminated
|
View Details
|
| Z-0044-2010 |
Manual Pulmonary Resuscitator, Adult w/ Large Adult Mask & Reservoir Bag. Product # 12-8705.
Packaged under the name of Evergreen.
Reason: In some Manual Pulmonary Resuscitator units the duckbill valve and retention ring inside the patient valve housing may become dislodged and fall into the MPR bag/body. If this occurs, the MPR may lose its ability to generate the positive pressure necessary necessary to ventilate the patient.
|
2009-10-06 |
Terminated
|
View Details
|