Trumpf Medical Systems, Inc.

Entity profile compiled from public source records

Public Record Profile Registrations: 0 510(k) filings: 0 Recalls: 4

Known Identifiers & Locations

FEI Numbers 3003184737
Owner Operator Numbers N/A
Locations Covered N/A

Registered Establishments (0)

No device registration and listing records found for this entity.

510(k) Premarket Notifications (0)

No 510(k) notifications found for this entity.

FDA Recall History (4)

Recall # Product Description Initiation Date Status Action
Z-0040-2017 Mobile Column TruSystem 7500, Material Number 1717023; Mobile Column TruSystem 5500, Material Number 1607822 Patient positioning during surgery, ranging from anesthesia induction to actual surgery to recover from the anesthesia. Also, patient transport on the operating table top from a patient transfer system to the operating theater of from the operating theater to a patient transfer system Reason: Three complaints describing TruSystem tables unexpectedly descending approximately 20 centimeters. 2016-07-10 Terminated View Details
Z-0041-2017 Stationary Column TruSystem 7599, Material Number 1717020 Patient positioning during surgery, ranging from anesthesia induction to actual surgery to recover from the anesthesia. Also, patient transport on the operating table top from a patient transfer system to the operating theater of from the operating theater to a patient transfer system. Reason: Three complaints describing TruSystem tables unexpectedly descending approximately 20 centimeters. 2016-07-10 Terminated View Details
Z-0042-2017 Floor Mounting Column TS 7500 U, Material Number 1730732; Floor Mounting Column ARTIS, Material Number 1501878 Patient positioning during surgery, ranging from anesthesia induction to actual surgery to recover from the anesthesia. Also, patient transport on the operating table top from a patient transfer system to the operating theater of from the operating theater to a patient transfer system. Reason: Three complaints describing TruSystem tables unexpectedly descending approximately 20 centimeters. 2016-07-10 Terminated View Details
Z-0043-2017 Operating Table TruSystem 7000 U, Material Number 1604788; Operating Table TruSystgem 7000 U (MB), Material Number 1604786 Patient positioning during surgery, ranging from anesthesia induction to actual surgery to recover from the anesthesia. Also, patient transport on the operating table top from a patient transfer system to the operating theater of from the operating theater to a patient transfer system Reason: Three complaints describing TruSystem tables unexpectedly descending approximately 20 centimeters. 2016-07-10 Terminated View Details

Profile Disclaimer

AvaSpecs compiles public data records from the FDA database. This compiled profile does not indicate endorsement, safety, clearance, approval, or compliance status. AvaSpecs is a private search index and is not affiliated with the FDA or any government agency.

Account Required Full available specs will require a free AvaSpecs account.