| Z-0015-2018 |
Philips Healthcare Brilliance 64 System, X-Ray, Tomography, Computed
These computed tomography x-ray systems are intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes.
Reason: During a bolus tracking procedure, no images were generated when the scan was completed, and the raw data file was not available for offline reconstruction. There is a discrepancy between the calculated reconstruction length and the actual scan length that results in the inability to reconstruct raw data. The operator may choose to rescan the patient.
|
2017-01-18 |
Terminated
|
View Details
|
| Z-0016-2018 |
Philips Healthcare Ingenuity Core X-Ray, Tomography, Computed
These computed tomography x-ray systems are intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes.
Reason: During a bolus tracking procedure, no images were generated when the scan was completed, and the raw data file was not available for offline reconstruction. There is a discrepancy between the calculated reconstruction length and the actual scan length that results in the inability to reconstruct raw data. The operator may choose to rescan the patient.
|
2017-01-18 |
Terminated
|
View Details
|
| Z-0017-2018 |
Philips Healthcare Ingenuity CT X-Ray, Tomography, Computed
These computed tomography x-ray systems are intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes.
Reason: During a bolus tracking procedure, no images were generated when the scan was completed, and the raw data file was not available for offline reconstruction. There is a discrepancy between the calculated reconstruction length and the actual scan length that results in the inability to reconstruct raw data. The operator may choose to rescan the patient.
|
2017-01-18 |
Terminated
|
View Details
|
| Z-0018-2018 |
Philips Healthcare Ingenuity Core128 X-Ray, Tomography, Computed
These computed tomography x-ray systems are intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes.
Reason: During a bolus tracking procedure, no images were generated when the scan was completed, and the raw data file was not available for offline reconstruction. There is a discrepancy between the calculated reconstruction length and the actual scan length that results in the inability to reconstruct raw data. The operator may choose to rescan the patient.
|
2017-01-18 |
Terminated
|
View Details
|
| Z-0019-2018 |
Philips Healthcare Brilliance iCT X-Ray, Tomography, Computed
These computed tomography x-ray systems are intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes.
Reason: During a bolus tracking procedure, no images were generated when the scan was completed, and the raw data file was not available for offline reconstruction. There is a discrepancy between the calculated reconstruction length and the actual scan length that results in the inability to reconstruct raw data. The operator may choose to rescan the patient.
|
2017-01-18 |
Terminated
|
View Details
|
| Z-0020-2018 |
Philips Healthcare Brilliance iCT SP X-Ray, Tomography, Computed
These computed tomography x-ray systems are intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes.
Reason: During a bolus tracking procedure, no images were generated when the scan was completed, and the raw data file was not available for offline reconstruction. There is a discrepancy between the calculated reconstruction length and the actual scan length that results in the inability to reconstruct raw data. The operator may choose to rescan the patient.
|
2017-01-18 |
Terminated
|
View Details
|