| Z-0019-2025 |
Tablo Console, REF: PN-0003000, PN-0006000, part of the Tablo Hemodialysis System. The Tablo Hemodialysis System is indicated for use in patients with acute and/or chronic renal failure, with or without ultrafiltration, in an acute or chronic care facility.
Reason: Blood leak sensor, a component of hemodialysis device, may show decreased sensitivity, at the lowest level, specified as 0.35 mL/min blood leak rate, which could delay triggering the ALARM_DIALYZER_BLOOD_LEAK alarm, if blood loss, less-than 300ml, occurs, the patient may experience hypotension, fatigue, cramps, dizziness, headaches, nausea, and shortness of breath.
|
2024-08-01 |
Open, Classified
|
View Details
|