| Registration # | Establishment Name | Location | Establishment Type(s) | Action |
|---|---|---|---|---|
| 1220592 | HOBBS MEDICAL, INC. | Stafford Springs, CT, US | Manufacture Medical Device,Manufacture Medical Device for Another Party (Contract Manufacturer) | View Details |
| Recall # | Product Description | Initiation Date | Status | Action |
|---|---|---|---|---|
| Z-0010-2023 | Hobbs Medical Spray Catheter: The Mistifier: 2.6mm 260cm Disposable For endoscopy procedures Ref: 2190 Reason: Device label display an incorrect expiration date, extends the shelf life of the product. Affected product distributed in the United States from 3/25/2018 to 9/28/2020, expiration dates between 2022-11-01 and 2025-09-01 (Devices have a lot number ending in R, or are kits containing at least 1 device with a lot number ending in R.) | 2022-07-29 | Open, Classified | View Details |
| Z-0011-2023 | Hobbs Medical Spray Catheter: The Mistifier: 2.6mm 165cm Disposable For endoscopy procedures Ref: 2191 Reason: Device label display an incorrect expiration date, extends the shelf life of the product. Affected product distributed in the United States from 3/25/2018 to 9/28/2020, expiration dates between 2022-11-01 and 2025-09-01 (Devices have a lot number ending in R, or are kits containing at least 1 device with a lot number ending in R.) | 2022-07-29 | Open, Classified | View Details |
| Z-0012-2023 | Hobbs Medical Flex-Ez Guide Wire Disposable For endoscopy procedures Ref: 3412 Reason: Device label display an incorrect expiration date, extends the shelf life of the product. Affected product distributed in the United States from 3/25/2018 to 9/28/2020, expiration dates between 2022-11-01 and 2025-09-01 (Devices have a lot number ending in R, or are kits containing at least 1 device with a lot number ending in R.) | 2022-07-29 | Open, Classified | View Details |
| Z-0013-2023 | Hobbs Medical Cytology Brush Catalog Number: 4204 Reason: Device label display an incorrect expiration date, extends the shelf life of the product. Affected product distributed in the United States from 3/25/2018 to 9/28/2020, expiration dates between 2022-11-01 and 2025-09-01 (Devices have a lot number ending in R, or are kits containing at least 1 device with a lot number ending in R.) | 2022-07-29 | Open, Classified | View Details |
| Z-0014-2023 | Hobbs Medical Cytology Brush Catalog Number: 4206 Reason: Device label display an incorrect expiration date, extends the shelf life of the product. Affected product distributed in the United States from 3/25/2018 to 9/28/2020, expiration dates between 2022-11-01 and 2025-09-01 (Devices have a lot number ending in R, or are kits containing at least 1 device with a lot number ending in R.) | 2022-07-29 | Open, Classified | View Details |
| Z-0015-2023 | Hobbs Medical Polypectomy Snare Catalog Number: 4563 Reason: Device label display an incorrect expiration date, extends the shelf life of the product. Affected product distributed in the United States from 3/25/2018 to 9/28/2020, expiration dates between 2022-11-01 and 2025-09-01 (Devices have a lot number ending in R, or are kits containing at least 1 device with a lot number ending in R.) | 2022-07-29 | Open, Classified | View Details |
| Z-0016-2023 | Hobbs Medical Polypectomy Snare Catalog Number: 4564 Reason: Device label display an incorrect expiration date, extends the shelf life of the product. Affected product distributed in the United States from 3/25/2018 to 9/28/2020, expiration dates between 2022-11-01 and 2025-09-01 (Devices have a lot number ending in R, or are kits containing at least 1 device with a lot number ending in R.) | 2022-07-29 | Open, Classified | View Details |
| Z-0017-2023 | Hobbs Transbronchial Aspiration Needle and Vacu-lok Aspirating Syringe Catalog Number: 4620 Reason: Device label display an incorrect expiration date, extends the shelf life of the product. Affected product distributed in the United States from 3/25/2018 to 9/28/2020, expiration dates between 2022-11-01 and 2025-09-01 (Devices have a lot number ending in R, or are kits containing at least 1 device with a lot number ending in R.) | 2022-07-29 | Open, Classified | View Details |
| Z-0018-2023 | Hobbs Transbronchial Aspiration Needle Catalog Number: 4621 Reason: Device label display an incorrect expiration date, extends the shelf life of the product. Affected product distributed in the United States from 3/25/2018 to 9/28/2020, expiration dates between 2022-11-01 and 2025-09-01 (Devices have a lot number ending in R, or are kits containing at least 1 device with a lot number ending in R.) | 2022-07-29 | Open, Classified | View Details |
| Z-0019-2023 | Hobbs Scope Introducer Catalog Number: 4650 Reason: Device label display an incorrect expiration date, extends the shelf life of the product. Affected product distributed in the United States from 3/25/2018 to 9/28/2020, expiration dates between 2022-11-01 and 2025-09-01 (Devices have a lot number ending in R, or are kits containing at least 1 device with a lot number ending in R.) | 2022-07-29 | Open, Classified | View Details |
| Z-0020-2023 | Hobbs Posi-Stop Injection Needle Catalog Number: 4702 Reason: Device label display an incorrect expiration date, extends the shelf life of the product. Affected product distributed in the United States from 3/25/2018 to 9/28/2020, expiration dates between 2022-11-01 and 2025-09-01 (Devices have a lot number ending in R, or are kits containing at least 1 device with a lot number ending in R.) | 2022-07-29 | Open, Classified | View Details |
| Z-0021-2023 | Hobbs Posi-Stop Injection Needle Catalog Number: 4704 Reason: Device label display an incorrect expiration date, extends the shelf life of the product. Affected product distributed in the United States from 3/25/2018 to 9/28/2020, expiration dates between 2022-11-01 and 2025-09-01 (Devices have a lot number ending in R, or are kits containing at least 1 device with a lot number ending in R.) | 2022-07-29 | Open, Classified | View Details |
| Z-0022-2023 | Hobbs Posi-Stop Injection Needle Catalog Number: 4712 Reason: Device label display an incorrect expiration date, extends the shelf life of the product. Affected product distributed in the United States from 3/25/2018 to 9/28/2020, expiration dates between 2022-11-01 and 2025-09-01 (Devices have a lot number ending in R, or are kits containing at least 1 device with a lot number ending in R.) | 2022-07-29 | Open, Classified | View Details |
| Z-0023-2023 | Hobbs Posi-Stop Injection Needle Catalog Number: 4718 Reason: Device label display an incorrect expiration date, extends the shelf life of the product. Affected product distributed in the United States from 3/25/2018 to 9/28/2020, expiration dates between 2022-11-01 and 2025-09-01 (Devices have a lot number ending in R, or are kits containing at least 1 device with a lot number ending in R.) | 2022-07-29 | Open, Classified | View Details |
| Z-0024-2023 | Hobbs Posi-Stop Injection Needle Catalog Number: 4720 Reason: Device label display an incorrect expiration date, extends the shelf life of the product. Affected product distributed in the United States from 3/25/2018 to 9/28/2020, expiration dates between 2022-11-01 and 2025-09-01 (Devices have a lot number ending in R, or are kits containing at least 1 device with a lot number ending in R.) | 2022-07-29 | Open, Classified | View Details |
| Z-0025-2023 | Hobbs Posi-Stop Injection Needle Catalog Number: 4722 Reason: Device label display an incorrect expiration date, extends the shelf life of the product. Affected product distributed in the United States from 3/25/2018 to 9/28/2020, expiration dates between 2022-11-01 and 2025-09-01 (Devices have a lot number ending in R, or are kits containing at least 1 device with a lot number ending in R.) | 2022-07-29 | Open, Classified | View Details |
| Z-0026-2023 | Hobbs Helical Retrieval Basket Catalog Number: 4806 Reason: Device label display an incorrect expiration date, extends the shelf life of the product. Affected product distributed in the United States from 3/25/2018 to 9/28/2020, expiration dates between 2022-11-01 and 2025-09-01 (Devices have a lot number ending in R, or are kits containing at least 1 device with a lot number ending in R.) | 2022-07-29 | Open, Classified | View Details |
| Z-0027-2023 | Hobbs Helical Retrieval Basket Catalog Number: 4808 Reason: Device label display an incorrect expiration date, extends the shelf life of the product. Affected product distributed in the United States from 3/25/2018 to 9/28/2020, expiration dates between 2022-11-01 and 2025-09-01 (Devices have a lot number ending in R, or are kits containing at least 1 device with a lot number ending in R.) | 2022-07-29 | Open, Classified | View Details |
| Z-0028-2023 | Hobbs Helical Retrieval Basket Catalog Number: 4814 Reason: Device label display an incorrect expiration date, extends the shelf life of the product. Affected product distributed in the United States from 3/25/2018 to 9/28/2020, expiration dates between 2022-11-01 and 2025-09-01 (Devices have a lot number ending in R, or are kits containing at least 1 device with a lot number ending in R.) | 2022-07-29 | Open, Classified | View Details |
| Z-0029-2023 | Hobbs Helical Retrieval Basket Catalog Number: 4816 Reason: Device label display an incorrect expiration date, extends the shelf life of the product. Affected product distributed in the United States from 3/25/2018 to 9/28/2020, expiration dates between 2022-11-01 and 2025-09-01 (Devices have a lot number ending in R, or are kits containing at least 1 device with a lot number ending in R.) | 2022-07-29 | Open, Classified | View Details |
| Z-0030-2023 | Hobbs Helical Retrieval Basket Catalog Number: 4822 Reason: Device label display an incorrect expiration date, extends the shelf life of the product. Affected product distributed in the United States from 3/25/2018 to 9/28/2020, expiration dates between 2022-11-01 and 2025-09-01 (Devices have a lot number ending in R, or are kits containing at least 1 device with a lot number ending in R.) | 2022-07-29 | Open, Classified | View Details |
| Z-0031-2023 | Hobbs Helical Retrieval Basket Catalog Number: 4824 Reason: Device label display an incorrect expiration date, extends the shelf life of the product. Affected product distributed in the United States from 3/25/2018 to 9/28/2020, expiration dates between 2022-11-01 and 2025-09-01 (Devices have a lot number ending in R, or are kits containing at least 1 device with a lot number ending in R.) | 2022-07-29 | Open, Classified | View Details |
| Z-0032-2023 | Hobbs Helical Retrieval Basket Catalog Number: 4830 Reason: Device label display an incorrect expiration date, extends the shelf life of the product. Affected product distributed in the United States from 3/25/2018 to 9/28/2020, expiration dates between 2022-11-01 and 2025-09-01 (Devices have a lot number ending in R, or are kits containing at least 1 device with a lot number ending in R.) | 2022-07-29 | Open, Classified | View Details |
| Z-0033-2023 | Hobbs Helical Retrieval Basket Catalog Number: 4832 Reason: Device label display an incorrect expiration date, extends the shelf life of the product. Affected product distributed in the United States from 3/25/2018 to 9/28/2020, expiration dates between 2022-11-01 and 2025-09-01 (Devices have a lot number ending in R, or are kits containing at least 1 device with a lot number ending in R.) | 2022-07-29 | Open, Classified | View Details |
| Z-0034-2023 | Hobbs Medical Grasping Forceps Catalog Number:5004 Reason: Device label display an incorrect expiration date, extends the shelf life of the product. Affected product distributed in the United States from 3/25/2018 to 9/28/2020, expiration dates between 2022-11-01 and 2025-09-01 (Devices have a lot number ending in R, or are kits containing at least 1 device with a lot number ending in R.) | 2022-07-29 | Open, Classified | View Details |
| Z-0035-2023 | Hobbs Medical Grasping Forceps Catalog Number: 5008 Reason: Device label display an incorrect expiration date, extends the shelf life of the product. Affected product distributed in the United States from 3/25/2018 to 9/28/2020, expiration dates between 2022-11-01 and 2025-09-01 (Devices have a lot number ending in R, or are kits containing at least 1 device with a lot number ending in R.) | 2022-07-29 | Open, Classified | View Details |
| Z-0036-2023 | Hobbs Retrieval Forceps Catalog Number: 5018 Reason: Device label display an incorrect expiration date, extends the shelf life of the product. Affected product distributed in the United States from 3/25/2018 to 9/28/2020, expiration dates between 2022-11-01 and 2025-09-01 (Devices have a lot number ending in R, or are kits containing at least 1 device with a lot number ending in R.) | 2022-07-29 | Open, Classified | View Details |
| Z-0037-2023 | Hobbs Biliary Pigtail Stent Kits (containing 3412) Catalog Number: 6001 Reason: Device label display an incorrect expiration date, extends the shelf life of the product. Affected product distributed in the United States from 3/25/2018 to 9/28/2020, expiration dates between 2022-11-01 and 2025-09-01 (Devices have a lot number ending in R, or are kits containing at least 1 device with a lot number ending in R.) | 2022-07-29 | Open, Classified | View Details |
| Z-0038-2023 | Hobbs Biliary Pigtail Stent Kits (containing 3412) Catalog Number:6004 Reason: Device label display an incorrect expiration date, extends the shelf life of the product. Affected product distributed in the United States from 3/25/2018 to 9/28/2020, expiration dates between 2022-11-01 and 2025-09-01 (Devices have a lot number ending in R, or are kits containing at least 1 device with a lot number ending in R.) | 2022-07-29 | Open, Classified | View Details |
| Z-0039-2023 | Hobbs Biliary Pigtail Stent Kits (containing 6105) Catalog Number: 6023 Reason: Device label display an incorrect expiration date, extends the shelf life of the product. Affected product distributed in the United States from 3/25/2018 to 9/28/2020, expiration dates between 2022-11-01 and 2025-09-01 (Devices have a lot number ending in R, or are kits containing at least 1 device with a lot number ending in R.) | 2022-07-29 | Open, Classified | View Details |
| Z-0040-2023 | Hobbs Biliary Pigtail Stent Kits (containing 3412) Catalog Number: 6024 Reason: Device label display an incorrect expiration date, extends the shelf life of the product. Affected product distributed in the United States from 3/25/2018 to 9/28/2020, expiration dates between 2022-11-01 and 2025-09-01 (Devices have a lot number ending in R, or are kits containing at least 1 device with a lot number ending in R.) | 2022-07-29 | Open, Classified | View Details |
| Z-0041-2023 | Hobbs Biliary Pigtail Stent Kits (contains 3412) Catalog Number: 6027 Reason: Device label display an incorrect expiration date, extends the shelf life of the product. Affected product distributed in the United States from 3/25/2018 to 9/28/2020, expiration dates between 2022-11-01 and 2025-09-01 (Devices have a lot number ending in R, or are kits containing at least 1 device with a lot number ending in R.) | 2022-07-29 | Open, Classified | View Details |
| Z-0042-2023 | Hobbs Biliary Pigtail Stent Kits (containing 3412) Catalog Number: 6028 Reason: Device label display an incorrect expiration date, extends the shelf life of the product. Affected product distributed in the United States from 3/25/2018 to 9/28/2020, expiration dates between 2022-11-01 and 2025-09-01 (Devices have a lot number ending in R, or are kits containing at least 1 device with a lot number ending in R.) | 2022-07-29 | Open, Classified | View Details |
| Z-0043-2023 | Hobbs Biliary Pigtail Stent Kits Catalog Number: 6104 Reason: Device label display an incorrect expiration date, extends the shelf life of the product. Affected product distributed in the United States from 3/25/2018 to 9/28/2020, expiration dates between 2022-11-01 and 2025-09-01 (Devices have a lot number ending in R, or are kits containing at least 1 device with a lot number ending in R.) | 2022-07-29 | Open, Classified | View Details |
| Z-0044-2023 | Hobbs Biliary Pigtail Stent Kits Catalog Number: 6105 Reason: Device label display an incorrect expiration date, extends the shelf life of the product. Affected product distributed in the United States from 3/25/2018 to 9/28/2020, expiration dates between 2022-11-01 and 2025-09-01 (Devices have a lot number ending in R, or are kits containing at least 1 device with a lot number ending in R.) | 2022-07-29 | Open, Classified | View Details |
AvaSpecs compiles public data records from the FDA database. This compiled profile does not indicate endorsement, safety, clearance, approval, or compliance status. AvaSpecs is a private search index and is not affiliated with the FDA or any government agency.