| Z-0006-2008 |
GE Disposable Multi Absorber Canister used with EZchange Module for the Aisys, Avance, Aespire Anesthesia systems. The EZchange Module is an optional accessory for these anesthesia systems.
Reason: Elevated inspired CO2 levels related to the use of the Datex-Ohmeda GE Healthcare Disposable Multi Absorber Canister with the EZchange Module.
|
2007-07-24 |
Terminated
|
View Details
|
| Z-0020-2009 |
GE Datex-Ohmeda Aisys, Datex-Ohmeda, Inc., a General Electric Company going to market as GE Healthcare, GE Healthcare PO Box 7550, Madison, WI. 53707-7550
Datex-Ohmeda Aisys
Intended to provide general inhalation anesthesia and ventilatory support to a wide range of patients. The device is intended for volume or pressure control ventilation. The Aisys is not suitable for use in a MRI environment.
Reason: The electronic vaporization system in the GE Aisys Anesthesia machine contains the following components: Backpressure Valve, Inflow Check Valve and Cassette Interface Board, that can contribute to independent failures.
|
2008-08-28 |
Terminated
|
View Details
|
| Z-0021-2007 |
GE Healthcare-Datex-Ohmeda Aisys Anesthesia Delivery System with Aladin 2 cassettes, GE Healthcare, Madison, WI; Datex-Ohmeda, inc., a General Electric company
Reason: Vaporizer shutdown with alarm-An intermittent leak in the Aladin 2 cassette/Aisys pneumatic interface (valves) may occur that will invoke an alarm state. When this occurs 'Vaporizer Failure' will appear on the Aisys main display, with an accompanying audible alrm. Delivery of anesthetic Agent is halted; gas flow and ventilation are not affected.
|
2006-08-10 |
Terminated
|
View Details
|
| Z-0037-06 |
Datex-Ohmeda S/5 Aespire Anesthesia System 7100
Reason: Two conditions may occur with the S/5 Aespire Anesthesia System. The internal auxiliary flowmeter tubing may become disconnected if the auxiliary oxygen supply tubing is occluded and the patient may not receive desired supplemental oxygen which could lead to serious injury if undetected. Also the O2 switch may leak or fail.
|
2005-09-27 |
Terminated
|
View Details
|
| Z-0038-06 |
Datex-Ohmeda S/5 Aespire Anesthesia System 7900
Reason: Two conditions may occur with the S/5 Aespire Anesthesia System. The internal auxiliary flowmeter tubing may become disconnected if the auxiliary oxygen supply tubing is occluded and the patient may not receive desired supplemental oxygen which could lead to serious injury if undetected. Also the O2 switch may leak or fail.
|
2005-09-27 |
Terminated
|
View Details
|
| Z-0039-06 |
Datex-Ohmeda Aespire Carestation. This device is not marketed in the US.
Reason: Two conditions may occur with the S/5 Aespire Anesthesia System. The internal auxiliary flowmeter tubing may become disconnected if the auxiliary oxygen supply tubing is occluded and the patient may not receive desired supplemental oxygen which could lead to serious injury if undetected. Also the O2 switch may leak or fail.
|
2005-09-27 |
Terminated
|
View Details
|