Known Identifiers & Locations
FEI Numbers
1419887
Owner Operator Numbers
N/A
Locations Covered
N/A
Registered Establishments (0)
No device registration and listing records found for this entity.
510(k) Premarket Notifications (0)
No 510(k) notifications found for this entity.
FDA Recall History (1)
| Recall # |
Product Description |
Initiation Date |
Status |
Action |
| Z-0001-2008 |
Bio-logic System Corp --Telescoping Camera Pole Assembly (part number 585-CYBPSM and 580-VIDPOL) and assemblies (580IPCART and 580 PNCART; Bio-Logic Systems Corporation, Mundelein, IL 60060; The Model 580-VIDPOL is used with the 580-PNCART that is wired for use with a Panasonic Camera. The pole is also a subassembly of the 585-CYBPSM with wires used in the 580-IPCART wired for the use with an Ipela Camera. Both carts are a component of the Ceegraph/Sleepscan Netlink with Video.
Reason: Bio-logic System Corp Camera Pole may weaken and detach at the base of the telescoping camera pole during transport causing pole to break at the welded base and fall.
|
2007-07-26 |
Terminated
|
View Details
|
Profile Disclaimer