| Z-0037-2024 |
API3200MD Mass Spectrometer, Part Number 4466230
Reason: In rare occasions, the combination of a leak of a highly flammable solvent at the ion source probe, and an electrical discharge may result in a flame on top of the probe and possible ejection of the electrode assembly from the probe. There is a risk to the patient/user, customer property, or environment because there is a potential for a health risk and property damage should the flame be left unattended. The failure causes a delay in patient results.
|
2023-08-30 |
Open, Classified
|
View Details
|
| Z-0038-2024 |
4500MD Triple Quad Mass Spectrometer, Part Number 5032522
Reason: In rare occasions, the combination of a leak of a highly flammable solvent at the ion source probe, and an electrical discharge may result in a flame on top of the probe and possible ejection of the electrode assembly from the probe. There is a risk to the patient/user, customer property, or environment because there is a potential for a health risk and property damage should the flame be left unattended. The failure causes a delay in patient results.
|
2023-08-30 |
Open, Classified
|
View Details
|
| Z-0039-2024 |
Citrine QTRAP Mass Spectrometer, Part Number 5063685
Reason: In rare occasions, the combination of a leak of a highly flammable solvent at the ion source probe, and an electrical discharge may result in a flame on top of the probe and possible ejection of the electrode assembly from the probe. There is a risk to the patient/user, customer property, or environment because there is a potential for a health risk and property damage should the flame be left unattended. The failure causes a delay in patient results.
|
2023-08-30 |
Open, Classified
|
View Details
|
| Z-0040-2024 |
Citrine Triple Quad Mass Spectrometer, Part Number 5063684
Reason: In rare occasions, the combination of a leak of a highly flammable solvent at the ion source probe, and an electrical discharge may result in a flame on top of the probe and possible ejection of the electrode assembly from the probe. There is a risk to the patient/user, customer property, or environment because there is a potential for a health risk and property damage should the flame be left unattended. The failure causes a delay in patient results.
|
2023-08-30 |
Open, Classified
|
View Details
|